Possibilities and analytical properties of the radiofrequency plasma pencil operating in the continuous and in the pulsed mode

Investor logo
Investor logo

Warning

This publication doesn't include Institute of Computer Science. It includes Central European Institute of Technology. Official publication website can be found on muni.cz.
Authors

NOVOSÁD Lukáš HRDLIČKA Aleš SLAVÍČEK Pavel OTRUBA Vítězslav KANICKÝ Viktor

Year of publication 2015
Type Article in Periodical
Magazine / Source Journal of analytical atomic spectrometry
MU Faculty or unit

Central European Institute of Technology

Citation
Web odkaz na článek
Doi http://dx.doi.org/10.1039/c4ja00402g
Field Analytic chemistry
Keywords plasma pencil; discharge; concentration; intensity; pulsed; continuous; mode; analysis; sample;
Description A two-electrode capacitively coupled plasma jet discharge operating at a 13.56 MHz symmetrical sinusoidal frequency carrier wave in a continuous mode and in a pulsed mode is described in this paper. The Internally pulsed mode was realised by the frequency modulation using a symmetrical rectangular frequency of 22 kHz with a duty cycle of 90%. The pulsed mode has an effect on the electron number densities ranging from 1 x 10(14) to 1.2 x 10(15) in the pulsed mode and from 1.7 x 10(14) to 1.0 x 10(15) cm(-3) in the continuous mode, gas temperatures 670-950 K in the pulsed and 630-880 K in the continuous mode, excitation temperatures 3100-4250 K in the pulsed and 3500-4280 K in the continuous mode, slopes and intercepts of the calibration lines and limits of detection and a weak effect on the coefficients of determination. All these parameters were measured at three different positions from the sample inlet of the plasma pencil. Most line intensities and the excitation temperature decreased while calcium lines and gas temperature calculated from OH radicals increased in the pulsed mode. These results indicate a strong interaction between calcium species and OH radicals. Analysis of real low mineralized water samples showed similarity of the yielded results to those obtained by reference ICP-OES.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.

More info