Rapid Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry Imaging with Scanning Desorption Laser Beam
Autoři | |
---|---|
Rok publikování | 2014 |
Druh | Článek v odborném periodiku |
Časopis / Zdroj | Analytical Chemistry |
Fakulta / Pracoviště MU | |
Citace | |
www | http://pubs.acs.org/doi/abs/10.1021/ac402823n |
Doi | http://dx.doi.org/10.1021/ac402823n |
Obor | Analytická chemie, separace |
Klíčová slova | Mass spectrometry imaging; laser beam scanning; MALDI; TOF; high-throughput |
Popis | This work shows that a substantial increase in the throughput of the TOF MS-based tissue imaging can be achieved by incorporating a mirror providing high-speed precision scanning of the laser beam along the sample surface. Equipped with the scanning mirror, a laboratory-built axial MALDI TOF MS instrument utilizing a 4-kHz UV laser recorded a 100 x 100 pixel MS image in 11 min using 100 laser shots per pixel. This is almost an order of magnitude faster when compared to a modern commercial instrument equipped with 1-kHz laser. |
Související projekty: |