Title in English Determination of iodine in milk by vapour-generation ICP-OES method
Authors

NIEDOBOVÁ Eva MACHÁT Jiří OTRUBA Vítězslav KANICKÝ Viktor

Year of publication 2005
Type Article in Proceedings
Conference Mikroelementy 2005
MU Faculty or unit

Faculty of Science

Citation
Field Analytic chemistry
Keywords iodine; milk; vapour generation; ICP-OES
Description In this work the method for determination of iodine in milk samples is descibed. The method for sample preparation was alkline ashing and iodine contents were measured with use of VG-ICP-OES. There are discussed spectral interferences and methods for its elimination in this work. Evaluated method was validated by analysis of CRM sample and by other independent technique ICP-MS.